1 Departments of Obstetrics and Gynecology, and Preventive Medicine, University of Southern California, Keck School of Medicine, Women's and Children's Hospital, Los Angeles, California; 2 Division of Endocrinology, Clinical Nutrition, and Vascular Medicine, Department of Internal Medicine, University of California at Davis, Sacramento, California; and 3 Clinical Research, Cancer Center, University of Virginia, Charlottesville, Virginia
Requests for reprints: Frank Z. Stanczyk, Departments of Obstetrics and Gynecology, and Preventive Medicine, University of Southern California, Keck School of Medicine, Women's and Children's Hospital, Room 1M2, 1240 North Mission Road, Los Angeles, CA 90033. Phone: 323-226-3220; Fax: 323-226-2850; E-mail: fstanczyk{at}socal.rr.com
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
|
These discrepancies in assay performance particularly limit investigations where comparisons of absolute, rather than relative, values of sex hormone concentrations are needed. Currently, estimating disease risk in relation to serum sex steroids relies on comparing relative risks of disease across tertiles or quartiles of sex steroid concentrations. This epidemiologic approach relies less on an assay producing accurate measurements of absolute concentration values. A sex steroid measurement will likely be classified into the correct tertile or quartile category as long as it is not near the border of two categories, thereby helping to circumvent assay inaccuracies. However, direct comparisons of sex hormone data across epidemiologic studies and recommendations about the use of hormone assays in the clinical setting would be facilitated by the use of assays across studies that provide both accurate and precise measurements. Determining accurate absolute concentration values becomes essential, for example, to examine dose-response relationships with exogenous hormone use or effects of clinically relevant hormonal threshold levels for disease outcomes. Current assay limitations potentially hinder further understanding of the biological significance of serum sex steroids and the ability to make clinical recommendations about hormonal strategies for prevention and treatment, based on actual concentration values. Furthermore, misleading information may be obtained if no attempt is made to determine how well assay measurements correlate against those obtained by a gold standard assay. This is especially true when commercial kits are used for measuring steroid hormones in epidemiologic studies. Assays done with such kits are rarely validated thoroughly by the kit manufacturer.
The purpose of this commentary is to (a) point out the current limitations of steroid hormone assays using serum or plasma, particularly in measurements of low concentrations of endogenous hormones (e.g., E2 or testosterone in women and E2 in men); (b) provide a perspective as to potential approaches to assay standardization; and (c) project when standardization of such assays may be accomplished.
| Immunoassay Methods |
|---|
|
|
|---|
The first steroid RIA method, which was developed by Abraham to quantify circulating levels of E2 (11, 12), consisted of purification of E2 in serum or plasma samples by organic solvent extraction and column chromatography, before its quantification by RIA. A small amount of tritiated E2 was added to the samples first to correct for procedural losses. Organic solvent extraction removed the unconjugated steroids, leaving behind all water-soluble conjugated steroids (sulfates and glucuronides). Conjugated steroids are present in considerably higher concentrations than unconjugated steroids in blood. Thus, conjugated steroids such as estrone sulfate, dehydroepiandrosterone sulfate, and androsterone sulfate can cross-react with the antibody used in an RIA and cause overestimation in the measurement of a steroid hormone. Column chromatography was used to separate E2 from E1 and other unconjugated steroids, which include numerous E2 metabolites. The E2 RIA method was validated and shown to be sensitive, specific, precise, and accurate.
The extraction/chromatographic RIA method developed for E2 was soon adapted to measure other steroid hormones. In addition, multiple analytes began to be measured in a relatively small sample volume. For example, androstenedione, DHEA, testosterone, E2, and E1 can be measured reliably in 0.6 mL of postmenopausal serum (13). In the last 35 years or so, the extraction/chromatographic RIA method has remained essentially the same and has been used in countless studies, yielding data that have enriched the field of endocrinology with new knowledge. In addition, the method has played an important role in measurements of a variety of steroid hormones used for diagnostic testing in clinical laboratories.
Direct Immunoassays
Although the extraction/chromatographic RIA method is highly reliable under most circumstances when thoroughly validated, it has some disadvantages. The method is cumbersome, time-consuming, and costly. For these reasons, direct RIAs, which do not incorporate a purification step before quantification of the analyte, were developed in the late 1970s. Kits that contained the necessary reagents to perform steroid RIAs became available commercially. Subsequently, the assay reagents were used in an instrument so that the assay could be done on an automated platform. This required replacement of the radioactive marker used in the assay with a nonradioactive one. Presently, direct immunoassay methods use chemiluminescent, fluorescent, or enzyme markers for quantification purposes.
Unlike extraction/chromatographic RIAs, which are time-consuming, direct immunoassay methods are simple, convenient, rapid, relatively inexpensive, and require less serum (usually only 0.1 mL). However, they can also have serious disadvantages. They often overestimate steroid concentrations due to lack of specificity of the antibody. For example, a major contributor to overestimation of levels in direct E2 and E1 immunoassays is estrone sulfate, which may cross-react substantially with E2 and E1 antisera used in immunoassays. This limitation is especially important in serum samples from postmenopausal women who have very low E2 and E1 levels but relatively high estrone sulfate levels. Also, matrix differences between the serum sample and solutions of standards may affect the results. This is particularly true of samples that are hemolyzed or lipemic. In addition, certain steroids (e.g., E2 and testosterone) may not be released efficiently from sex hormone–binding globulin with the specific reagent used by the kit manufacturer to accomplish this. This could underestimate the levels. A particularly important and largely unrecognized problem is that direct assays lack sufficient sensitivity and specificity to measure low levels of steroids such as E2, E1, and testosterone in serum from postmenopausal women and both estrogens in male serum.
Evidence showing that certain direct steroid immunoassays carried out with reagents in commercial kits are not reliable can be found in several studies. In our study, we evaluated four different commercially available direct testosterone immunoassay kits and used our extraction/chromatographic testosterone RIA as a standard for comparison in 10 premenopausal and 10 postmenopausal women, as well as 10 men (14). Our results show that the four assays using the kits did generally well for male serum samples, but samples from premenopausal and postmenopausal women showed poor accuracy as they significantly deviated (P < 0.05) from the standard values based on ANOVA testing and/or showed poor intraclass correlations (assay correlations in premenopausal samples: 0.20, 0.52, 0.70, 0.71, and in postmenopausal samples: 0.62, 0.64, 0.88, 0.95). Our findings are consistent with those reported by Taieb et al. (15) who measured serum testosterone levels in women, men, and children by use of 10 different direct testosterone immunoassay kits and by isotope dilution gas chromatography-mass spectrometry (GC-MS). Compared with GC-MS, 7 of the 10 direct immunoassays overestimated testosterone concentrations in samples from women. The mean testosterone values measured by the immunoassays were 46% above those obtained by GC-MS. In samples from men, the direct immunoassays underestimated testosterone concentrations, giving mean results that were 12% below those obtained by GC-MS. On the basis of their results, they concluded that the direct assays were acceptable for measuring testosterone in male samples, but not in samples from women or children. In an accompanying editorial (16) on the article by Taieb et al., the editors concluded that "guessing seems to be nearly as good as most commercially available immunoassays and clearly superior to some."
In the same study in which we evaluated the reliability of direct testosterone immunoassay kits, we also compared female serum E2 levels in 30 women determined by direct E2 immunoassay kits to those obtained by our extraction/chromatographic E2 RIA (14). Again, we found that the direct immunoassays were unreliable, as they deviated significantly (P < 0.05) from the standard values based on ANOVA testing and/or showed poor intraclass correlation (<0.85 and as low as 0.38). Our findings were substantiated by Nelson et al. (17), who showed that not only direct E2 immunoassays, but also direct E1 immunoassays were unreliable. For comparison, they used a liquid chromatography-tandem MS (LC-MS/MS) assay for simultaneous measurement of E2 and E1 in female plasma. At low E2 concentrations, two of the three direct immunoassays showed poor agreement and the third direct immunoassay showed no agreement with LC-MS/MS. E1 results obtained by direct RIA correlated moderately well with E1 measured by LC-MS/MS (adjusted r2 = 0.67). However, E1 measurements obtained by an RIA method that used an organic solvent extraction step before the RIA showed good to excellent agreement with LC-MS/MS.
In a pilot study of elderly postmenopausal women with low E2 levels (<20 pg/mL), we found that three extraction/chromatographic RIAs were highly correlated in their E2 measurements, whereas four direct immunoassays were not (18). Table 2 shows a comparison of the lower detection limits and required serum volumes among these assays to highlight the variability in assay performance and feasibility for research studies. The extraction/chromatographic RIAs resembled more closely the values obtained with a comparison method of GC-MS/MS and one another (correlations ranging from 0.82 to 0.94) than did the direct immunoassay kits (correlations ranging from 0.38 to 0.83). A confirmatory study compared the direct immunoassay and the extraction-based RIA, which correlated best with GC-MS/MS in the pilot study, to GC-MS/MS in a larger sample of 374 older postmenopausal women not taking hormone therapy (18). The direct immunoassay correlated significantly weaker (P < 0.01) with GC-MS/MS than did the extraction-based RIA (Table 2). Another recent study compared very low E2 concentrations using RIA with a preceding extraction step and two direct immunoassays in women who, as a result of taking an aromatase inhibitor, had very low circulating E2 levels (19). At least 70% of the E2 measured by the two direct assays were estimated to be an artifact. Furthermore, adding a preextraction step to the use of the direct assays led to measurements resembling those from the extraction-only RIA. In the latter study, it was concluded that the use of direct assays alone "is inappropriate and is likely to give aberrant guidelines."
|
-androstanediol glucuronide by direct immunoassay can provide valuable data in breast cancer or prostate cancer studies, due to the relatively high concentrations of these compounds in blood. Use of direct immunoassays for measuring those hormones has a big advantage in providing data simply and rapidly. If a conventional RIA is used to quantify the same hormones, it is essential to split (by hydrolysis) the sulfate or glucuronide group off the molecule and then measure the unconjugated moiety. Such an assay is very cumbersome because it is essential first to remove the unconjugated steroid in the sample by organic solvent extraction, and then to hydrolyze the sulfate or glucuronide group off the molecule, using an enzyme (for estrone sulfate and 3
-androstanediol glucuronide) or acidic conditions (for dehydroepiandrosterone sulfate). After that, another organic solvent extraction step is required, followed by chromatography and quantification by RIA. Due to the multiple steps that are involved in such assays, procedural losses are often high. In addition, because the unconjugated steroid is being measured after hydrolysis of the conjugated steroid, a correction is often made to account for the difference in molecular weight between the conjugated and unconjugated steroid. Thus, in this case, a properly validated direct immunoassay is preferred over the conventional RIA. However, the best method for measuring a conjugated steroid is by a MS assay, because the entire molecule can be measured with high specificity and precision. | MS Assay Methods |
|---|
|
|
|---|
There is a paucity of data comparing serum steroid hormone levels measured by RIA and MS assay methods. However, in the randomized clinical trial involving 374 postmenopausal women (18), we found that the extraction/chromatographic RIA [r = 0.50; 95% confidence interval (95% CI), 0.40-0.59] and GC-MS/MS (r = 0.52; 95% CI, 0.43-0.61) correlated better with body mass index (BMI), a biological indicator of estrogen, than the direct immunoassay (r = 0.29; 95% CI, 0.15-0.41; P = 0.01). The extraction/chromatographic RIA (r = 0.89; 95% CI, 0.81-0.93) correlated more closely with GC-MS/MS than did the direct immunoassay (r = 0.65; 95% CI, 0.52-0.76; P < 0.01). Consistent with these findings, the correlation between the direct immunoassay and the extraction/chromatographic RIA was 0.66 (95% CI, 0.54-0.75). The extraction/chromatographic RIA also correlated better on retest with GC-MS/MS than did the direct assay. Two commercial laboratories have shown that there is a good correlation between extraction/chromatographic RIAs and LC-MS/MS assays (20-22). However, steroid hormone levels obtained by GC-MS/MS or LC-MS/MS assays are generally lower than corresponding values achieved by RIAs, presumably due to the greater specificity of the former assays (20-23). New reference ranges will be required for steroid hormones measured by MS methods.
LC-MS/MS and GC-MS/MS methodologies have the capability for achieving not only high sensitivity and specificity, and measurement of multiple analytes in a small sample volume, but also high throughput of samples. MS assays are considered to become the gold standard for steroid hormone measurements. However, presently, the use of this method has been restricted to a relatively small number of laboratories. This is due to the high cost of the mass spectrometer, the requirement for a highly trained individual to operate the instrument, and the time required to develop assays for a variety of steroid hormones. Nevertheless, the initial cost of the instrumentation and assay development can be offset by the relatively rapid turnaround time of highly accurate steroid measurements, compared with the costly, cumbersome, and time-consuming extraction/chromatography RIA methodology.
Another very important advantage of MS assays is that, eventually, profiles of steroid hormones will be measured in a single aliquot of serum. Presently, this can be done with urinary estrogens. Recently, Xu et al. (24) developed a LC-MS/MS assay for measuring the absolute quantities of 15 endogenous estrogens in urine obtained from premenopausal or postmenopausal women and from men. The method requires a single hydrolysis/extraction/derivatization step before quantification by LC-MS/MS and uses only 0.5 mL of urine. Although this method is promising, it needs to be extended to serum for measurement of principal estrogens and androgens and their metabolites in women and men.
| Sensitivity of E2 Assay Methods |
|---|
|
|
|---|
To enhance assay sensitivity, in vitro recombinant DNA bioassays, with sensitivities ranging from 0.02 to 1 pg/mL, have been developed (27, 28). These assays detected 2- to 3-fold lower mean levels of E2 than the extraction/chromatographic RIA method done on the same samples in normal postmenopausal women. For example, E2 values were 1.95 ± 0.52 pg/mL by ultrasensitive yeast bioassay and 5.90 ± 1.40 pg/mL by extraction/chromatographic RIA in postmenopausal women with breast cancer (28). Also, the cell bioassay detected mean values of 3.23 ± 2.96 pg/mL versus 11.9 ± 12.0 by extraction-only/RIA in 30 normal postmenopausal women (28). However, the correlations between the E2 values with the two methods were excellent (r = 0.84, r = 0.79). It is not known whether these bioassays more truly reflect circulating values or are subject to artifactual influences that lower the levels detected. Notwithstanding this issue, the ultrasensitive bioassays are too cumbersome for routine use and are relegated to research laboratories for special tasks such as assessing small increments of E2 absorption from the vagina (29).
An important question then is whether extraction/chromatographic RIA or MS assay methods are more sensitive for measurement of E2 in the lowest quartile samples. One GC-MS/MS assay has a level of sensitivity of 0.6 pg/mL and detected E2 in all 30 samples from normal postmenopausal women and 45 samples from postmenopausal women with breast cancer (30). Our recent review examined seven different RIAs that had detection limits ranging from 0.8 to 5 pg/mL (Table 2; ref. 18). These observations suggest that MS methods might have superior operating levels of sensitivity for measurement of samples with low E2 levels, where enhanced sensitivity is important. Further study is needed to directly compare the sensitivities of various extraction/chromatographic RIA methods with both GC-MS/MS and LC-MS/MS assay methods in this subgroup of postmenopausal women.
| Standardization of Steroid Hormone Assays in Serum/Plasma |
|---|
|
|
|---|
Investigators planning epidemiologic studies with steroid hormone measurements would benefit from conducting run-in pilot tests before committing large samples to a chosen assay. Such pilot tests would evaluate a selected assay or several possible assays against more than one reference criterion, such as MS and BMI, for accuracy and reliability. However, this is not sufficient for ultimate accuracy because variation often occurs over time in the course of a study, which may ultimately compromise study results.
As stated earlier, there seems to be general agreement that MS assays will become the gold standard for steroid hormone measurements. However, currently, there is a misconception that whenever current MS methods are used to quantify a steroid hormone, the resulting value is very accurate and is obtained by the gold standard. It is important to realize that interlaboratory differences exist in steroid hormone concentrations measured by MS assays. This is due to the fact that conditions used to perform the assay may vary from laboratory to laboratory. Assay conditions that can affect the measurements include derivative preparation, type of internal standard, calibrator purity, and type of instrumentation. Thus, to standardize MS assays for measurement of steroid hormones, it is essential that assay conditions (reagents, instruments, and procedure) be the same in all laboratories involved in standardizing an assay. In addition, it is essential to minimize intra-assay and intralaboratory variability. Assay standardization will require a great effort by participating laboratories.
| Recommended Approach to Assay Standardization |
|---|
|
|
|---|
For validation of individual assays, whether using direct immunoassay, extraction-only RIA, extraction/chromatographic RIA, or MS assay methods, it will be necessary to determine agreed upon confidence limits related to measurements of different steroid hormones in standard pools. This criterion would allow validation of sensitivity, specificity, precision, and accuracy of current steroid hormone assay methodology and would provide a surrogate for a true gold standard. The Endocrine Society in conjunction with the Center for Disease Control is currently exploring this concept and should be strongly encouraged to establish such a program (31).
Received 9/10/06; revised 6/19/07; accepted 7/ 2/07.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. J. Soldin and O. P. Soldin Steroid Hormone Analysis by Tandem Mass Spectrometry Clin. Chem., June 1, 2009; 55(6): 1061 - 1066. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. W. Vesper and L. M. Thienpont Traceability in Laboratory Medicine Clin. Chem., June 1, 2009; 55(6): 1067 - 1075. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Fan, W. Yue, J.-P. Wang, S. Aiyar, Y. Li, T.-H. Kim, and R. J. Santen Mechanisms of Resistance to Structurally Diverse Antiestrogens Differ under Premenopausal and Postmenopausal Conditions: Evidence from in Vitro Breast Cancer Cell Models Endocrinology, May 1, 2009; 150(5): 2036 - 2045. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. T. Falk, X. Xu, L. Keefer, T. D. Veenstra, and R. G. Ziegler A Liquid Chromatography-Mass Spectrometry Method for the Simultaneous Measurement of 15 Urinary Estrogens and Estrogen Metabolites: Assay Reproducibility and Interindividual Variability Cancer Epidemiol. Biomarkers Prev., December 1, 2008; 17(12): 3411 - 3418. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. M. Thienpont, K. Van Uytfanghe, S. Blincko, C. S. Ramsay, H. Xie, R. C. Doss, B. G. Keevil, L. J. Owen, A. L. Rockwood, M. M. Kushnir, et al. State-of-the-Art of Serum Testosterone Measurement by Isotope Dilution-Liquid Chromatography- Tandem Mass Spectrometry Clin. Chem., August 1, 2008; 54(8): 1290 - 1297. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Cancer Research | Clinical Cancer Research |
| Cancer Epidemiology Biomarkers & Prevention | Molecular Cancer Therapeutics |
| Molecular Cancer Research | Cancer Prevention Research |
| Cancer Prevention Journals Portal | Cancer Reviews Online |
| Annual Meeting Education Book | Meeting Abstracts Online |