National Institute of Industrial Health, Kawasaki 214-8585, Japan [R. Y., Y. O.], and Institute of Industrial Ecological Science, University of Occupational and Environmental Health, Kitakyushu 807-8555, Japan [H. K.]
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Yin et al. (8) found a good correlation between ELISA and HPLC-ECD measurements of 8-oxo-dG in DNA, but immunoaffinity purification of hydrolysates was required before the ELISA. Cooke et al. (6) adapted a commercially available version of the ELISA kit to measure 8-oxo-dG in calf thymus DNA and obtained almost the same range of values as measured by HPLC, but they did not report any correlation coefficient or the results of a regression analysis. The ELISA kit used by Cooke et al. contains anti-8-oxo-dG monoclonal antibody N45.1, whose specificity for 8-oxo-dG has been confirmed (9 , 10) . Monoclonal antibody N45.1 has better specificity for 8-oxo-dG (tested for 8-oxo-dG/2'-deoxyguanosine, 8-oxo-dG/8-oxo-G, and 8-oxo-dG/8-oxo-7,8-dihydroguanine) than other 8-oxo-dG antibodies (11) . Although 8-oxo-G (oxidation of guanine in RNA at C-8 position) demonstrated some cross-reactivity with N45.1, it required 2 orders higher concentration than 8-oxo-dG (10) . Urinary 8-oxo-dG has been measured with this antibody in several studies (12 , 13) , but the correlation between the values obtained by ELISA and those obtained by HPLC-ECD was not acceptable (13) . Since then, the ELISA kit containing N45.1 has been improved, and a new version is now commercially available. Saito et al. (14) obtained acceptable results with the improved ELISA kit in measurements of urinary 8-oxo-dG by a recovery test in which several concentrations of 8-oxo-dG solution were added and by a dilution test in which urine samples were diluted severalfold with PBS. However, no results of comparisons between urinary 8-oxo-dG values measured by the improved ELISA and by HPLC-ECD have ever been reported.
Because measurement of urinary 8-oxo-dG has become a popular means of evaluating the level of oxidative stress in the human body, it seemed worthwhile to determine whether the improved ELISA kit is a good alternative for measuring urinary 8-oxo-dG. In the present study, we measured urinary 8-oxo-dG by HPLC-ECD and with the improved ELISA kit and compared these two urinary 8-oxo-dG measurements.
| Materials and Methods |
|---|
|
|
|---|
Measurement of 8-oxo-dG by HPLC-ECD.
The urinary 8-oxo-dG levels were determined as described previously (15)
. Briefly, the urine specimen was injected into the first HPLC column [ShodexAsahipak GS-320HQ 7G (500 x 7.6 mm); elution, 0.1% acetic acid]. An aliquot (100 µl) of the fraction containing 8-oxo-dG (5061 min) was automatically injected into the second HPLC column [YMC-Pack ODS-AM (250 x 4.6 mm); elution, aqueous methanol (5%) containing 35 mM NaOAc, 12.5 mM citric acid (pH 7.5)] by an autosampling injector (231 XL; Gilson), and 8-oxo-dG was measured by an EC detector (Coulochem II, ESA; guard cell, 350 mV; channel 1, 300 mV; channel 2, 150 mV).
Measurement of 8-oxo-dG by ELISA.
Thawed urine specimens were centrifuged at 4000 x g for 10 min, and the supernatant was used to measure 8-oxo-dG with a competitive ELISA kit named 8-OH-dG Check (Japan Institute for the Control of Aging, Fukuroi, Japan or Genox Corp., Baltimore, MD). The specificity of monoclonal antibody N45.1 used in the kit was confirmed in the previous study (9
, 10)
. All materials in the ELISA kit were kept at 4°C until the day of the experiment. A diagram of the ELISA method is shown in Fig. 1
. The plate layout used in our experiment is shown in Fig. 2
. According to the protocol provided by Japan Institute for the Control of Aging, the outermost wells were not used for measurement due to concerns about possible edge effects (Fig. 2)
; however, the same volume (50 µl) of water was added into the outermost wells to maintain the uniform temperature within the wells. All standard solutions, i.e., S1S6 (0.5, 2, 8, 20, 80, and 200 ng/ml), and the urine specimens were prepared in triplicate (Fig. 2)
. In blank well (line H), 50 µl of water were also added instead of samples, and the same procedure was performed before adding chromatic solution (Fig. 2)
. C1, C2, and C3 were urinary controls (high, medium, and low levels of 8-oxo-dG in urine, respectively) to confirm whether each experiment proceeded properly (Fig. 2)
. As a preliminary test, we measured 8-oxo-dG levels of the samples from eight volunteers with the ELISA kit to select the controls. The preliminary test was done three times on different days. The 8-oxo-dG levels of samples ranged from 3.0 to 23.3 ng/ml. We selected three samples, one with high 8-oxo-dG levels (about 20 ng/ml), another with medium levels (about 10 ng/ml), and the last with low levels (about 3 ng/ml), as controls C1, C2, and C3, respectively. Experiments were performed according to the instructions in the kit, except for the use of our three controls and changing plate layout slightly. After adding primary and secondary antibody, we incubated the plate at 37°C for 1 h using Incubator IC400 (Yamato, Japan). Incubator IC400 circulates warm air by thermal convection, and temperature was controlled at 37 ± 1°C. To maintain the same plate temperature for each experiment, we selected one location in the incubator for every experiment. The absorbances of the wells were measured at 450 nm using Wellreader sme 3400 (Iwaki, Japan). A calibration curve with concentrations of the standard 8-oxo-dG solutions (in logarithm) versus the absorbances (in linear) was constructed for each experiment and used to calculate the 8-oxo-dG concentrations in urine specimens.
|
|
| Results |
|---|
|
|
|---|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Prieme et al. (13) evaluated an ELISA including the same monoclonal antibody (N45.1) and reported 8-fold higher results on average as compared with HPLC-ECD. The correlation between the urinary 8-oxo-dG values obtained by HPLC-ECD and the ELISA was weak (r = 0.42; Ref. 13 ). The difference between our results and the results of the experiments performed by Prieme et al. (13) may be attributable to the use of different versions of the ELISA kit. We used the improved version of the kit, which differs somewhat from the old version used by Prieme et al. (product information from Japan Institute for the Control of Aging), for example, in the amount of 8-oxo-dG-protein conjugate coated on the plate, the concentration of standard solution (old, 0.642000 ng/ml; new, 0.5200 ng/ml), the concentration of the primary monoclonal antibody, the buffer contents, and the type of chromatic solution. Thus, changes in the quality of the kit may be one reason why we obtained a better correlation between the HPLC-ECD and the ELISA compared with the results of Prieme et al. (13) . Another reason for the difference may be our use of three controls with appropriate concentrations and our strict selection of reliable data. The study of Prieme et al. (13) did not mention the use of appropriate controls in their experiment.
We obtained a good correlation between the results of HPLC-ECD and the ELISA; however, the 8-oxo-dG values obtained by the ELISA were still nearly 2-fold higher than the values obtained by HPLC-ECD. Although the monoclonal antibody N45.1 has high specificity for 8-oxo-dG (9, 10, 11) , there may be other unknown substances in urine that react with the antibody. There is evidence that the same values were obtained by the ELISA and HPLC-ECD if urine was passed through the first HPLC-ECD column, and the 8-oxo-dG fraction was collected before the ELISA.3 Because all urine specimens had been centrifuged before the measurement by both methods, the first HPLC column might have removed some unknown fractions that were not removed by the pretreatment.
According to Toyokuni et al. (10) , N45.1 also reacted with 8-oxo-G (RNA form), although its reaction required 2 orders higher concentration of 8-oxo-G than that of 8-oxo-dG. Thus, a high concentration of urinary 8-oxo-G might interfere with the analysis. It has been reported that RNA was more sensitive to oxidative stress than DNA because it was easier to induce 8-oxo-G than 8-oxo-dG in livers of rats (16) and in human skin fibroblasts (17) . In addition, because it has been considered that turn over of RNA is faster than that of DNA (18) , there might be a large amount of 8-oxo-G in the urine. Existence of a urinary 8-oxo-G peak was confirmed, although its concentration was not compared with urinary 8-oxo-dG.4 Accordingly, urinary 8-oxo-G might react with N45.1. This is a possible reason why the ELISA method showed two times higher values than the HPLC-ECD method.
In the study performed by Saito et al. (14) , the urinary 8-oxo-dG levels obtained by the improved ELISA with plate incubation in air in uneven temperature condition were 2-fold higher than those obtained with plate incubation in water that better maintained the temperature. The 8-oxo-dG values they obtained with better temperature control were 12.0, 13.1, and 20.9 ng/ml, which were within our range of 8-oxo-dG values by the ELISA (2.0831.60 ng/ml). Thus, good temperature control is critical. Whereas the kit instructions suggest that the outermost wells should not be used, our data (data not shown), in which we measured the absorbance of the same sample in all wells, indicate that it is not a problem. In addition, strict selection of appropriate data using values in assay controls should also improve accuracy.
In view of the good correlation between the 8-oxo-dG values measured by HPLC-ECD and the ELISA, the ELISA can at least be applied to studies comparing relative urinary 8-oxo-dG values in molecular epidemiological studies to assess the risk of cancer or other diseases from environmental chemicals. The choice between HPLC-ECD and the ELISA may depend on the apparatus, analytical techniques, or manpower that each laboratory can afford. Although expensive instrumentation is needed for HPLC-ECD analysis, this method is recommended when more precise and accurate measurements are required.
In summary, we have demonstrated that the ELISA may be applicable to epidemiological studies because of its correlation with the HPLC-ECD method. However, further efforts to improve the assay accuracy and precision are needed.
| Footnotes |
|---|
1 To whom requests for reprints should be addressed, at National Institute of Industrial Health, 6-21-1 Nagao, Tama-ku, Kawasaki 214-8585, Japan. Phone: 044-865-6111; Fax: 044-865-6124; E-mail: ogawa{at}niih.go.jp ![]()
2 The abbreviations used are: 8-oxo-dG, 8-oxo-7,8-dihydro-2'-deoxyguanosine; HPLC-ECD, high-performance liquid chromatography with electrochemical detection; 8-oxo-G, 8-oxo-7,8-dihydro-2'-guanosine; CV, coefficient of variance. ![]()
3 Shimoi, K., Kasai, H., Yokata, N., Tokokoni, S., Kinae, N., Cancer Epidemiol Biomarkers Prev. 11, 2002. ![]()
Received 9/21/01; revised 5/17/02; accepted 5/29/02.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. S. Cooke, R. Olinski, S. Loft, and members of the European Standards Committee on Uri Measurement and Meaning of Oxidatively Modified DNA Lesions in Urine Cancer Epidemiol. Biomarkers Prev., January 1, 2008; 17(1): 3 - 14. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Peng, L.-Q. Li, M.-H. Peng, Z.-M. Liu, T.-W. Liu, Y. Guo, K.-Y. Xiao, Z. Qin, X.-P. Ye, X.-S. Mo, et al. Evaluation of oxidative stress in a group of adolescents exposed to a high level of aflatoxin B1 a multi-center and multi-biomarker study Carcinogenesis, November 1, 2007; 28(11): 2347 - 2354. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Yamauchi, H. Nakano, J. Maekawa, Y. Okamoto, Y. Ohnishi, T. Suzuki, and H. Kimura Oxidative Stress in Obstructive Sleep Apnea Chest, May 1, 2005; 127(5): 1674 - 1679. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. A. Hakim, R. B. Harris, H-H. S. Chow, M. Dean, S. Brown, and I. U. Ali Effect of a 4-Month Tea Intervention on Oxidative DNA Damage among Heavy Smokers: Role of Glutathione S-Transferase Genotypes Cancer Epidemiol. Biomarkers Prev., February 1, 2004; 13(2): 242 - 249. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Yoshida, Y. Ogawa, I. Mori, A. Nakata, R. Wang, S. Ueno, I. Shioji, and N. Hisanaga Associations between oxidative stress levels and total duration of engagement in jobs with exposure to fly ash among workers at municipal solid waste incinerators Mutagenesis, November 1, 2003; 18(6): 533 - 537. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. A. Hakim, R. B. Harris, S. Brown, H-H. S. Chow, S. Wiseman, S. Agarwal, and W. Talbot Effect of Increased Tea Consumption on Oxidative DNA Damage among Smokers: A Randomized Controlled Study J. Nutr., October 1, 2003; 133(10): 3303S - 3309. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Cancer Research | Clinical Cancer Research |
| Cancer Epidemiology Biomarkers & Prevention | Molecular Cancer Therapeutics |
| Molecular Cancer Research | Cancer Prevention Research |
| Cancer Prevention Journals Portal | Cancer Reviews Online |
| Annual Meeting Education Book | Cell Growth & Differentiation |